Central Schematic Diagram Mains and branches in piping systems shall be not less Outlet Stations
than 1/2'' nominal size.
O2 Pipeline
6 1/2'' MEDICAL GAS CONTROL VALVES Option 1 Copper
CLOSE IN EMERGENCY ONLY ZONE VALVE
FOR USE WITH 1/2" O.D. To Vac.
Automatic Manifold System
Oxygen
O 2 (PSI)
Nitrous Pipeline
Oxide
(PSI)
ZONE VALVE #
PULL RING FOR ACCESS Copper Control
N 2 O 1/2" O.D. Module
N2O Pipeline Copper
Wall Mount Cat 6 cable to 3/8" O.D. with
Bracket
Office Alarm Panel Zone Valve (100') Zone Valve Option 2 N2O O2 VAC
Outlet Requirements: N2O O2 VAC
Each outlet station for medical
gas with threaded (DISS) or
non-interchangeable quick Dual and Triple Micro Outlets are perfect companions
Copper to copper joints shall be braised using connect coupler shall be gas for flowmeter systems that are mounted in a cabinet,
copper-phosphorus or copper phosphorus-silver
Cat 6 cable brazing filler metal (B cup series) without flux with specific and shall consist of a like the Digital Ultra™ Flushmount Flowmeters and
to Office at least 1000˚ F. melting point. While being primary and secondary valve. the Ultra PC™ % Cabinet Mount Flowmeters. The
Alarm Panel brazed, joints shall be continually purged with Micros can rest on a shelf or the floor of a base
(100') oil-free dry nitrogen. The flow of purge gas shall cabinet and offer flexibility that is not always possible
be maintained until the joint is cool to touch. with a permanently mounted outlet station.
Dental Office Area Treatment Area
Cylinder Storage Area (Central System Manifold Room)
Cat 6 to
Alarm Panel
Cat 6 to 110 volt
Zone Valve outlet within 6’
Optional Source Valves
Pressure Relief Valve (N2O) Pressure Relief Valve (O2)
N2O DISS Hose O2 DISS Hose (3 feet)
(3 feet) O2 Regulator
N2O Regulator 5’ (60”) AFF
(Above
Finished
Nitrous Nitrous Floor) Oxygen Oxygen
Oxide Oxide
Cylinder Restraints
Nitrous Oxide Supply Oxygen Supply