Accutron™ Digital Ultra™ Flowmeter

           Our Accutron™ Digital Ultra™ Flowmeter features automatic compensation to easily adjust
           total flow or gas ratio while the other stays constant and both digital and electronic flowtube
           displays for ease of use . For patient safety, the maximum nitrous oxide ratio is 70% . Audible
           and visual alarms provide notification of low oxygen . For additional safety, these Flowmeters
           feature oxygen fail-safe and flush .


           Also includes emergency oxygen resuscitation connector as well as a sealed flush surface to
           make it easy to disinfect or barrier-protect .





           Dimensions Control Module:                           Oxygen Flow Range: 1 .0 to 9 .9 LPM
           7 .5 in x 7 in x 8 .75 in (19 .1 cm x 17 .8 cm x 22 .2 cm)  Nitrous Oxide Flow Range: 0 to 6 .9 LPM
           Weight: 6 .3 lbs                                     Inlet Pressure (each gas): 50 to 55 psi
           Mixed Gas Outlet Diameter:  .875 in Nominal (22mm)   Resuscitator Connector: Quick Connect ( .25 in Imperial)
           Oxygen Inlet: Male DISS (CGA 1240)                   Resuscitator Connector Flow: 100 LPM Oxygen
           Nitrous Oxide Inlet: Male DISS (CGA 1040)            Power Requirements: 100-240 V AC, 50-60 Hz, 30 Watts
           Nitrous Oxide Mixed Ratio Range: 0 to 70%